Accuracy of implant placement based on pre-surgical planning of three-dimensional cone-beam images: a pilot study
Penerbit
: Blackwell Synergy
Ringkasan :Aim: To evaluate the precision of transfer of a computer-based three-dimensional
(3D) planning, using re-formatted cone-beam images, for oral implant placement in
partially edentulous jaws.
Material and Methods: Four formalin-fixed cadaver jaws were imaged in a 3D
Accuitomo FPDs cone-beam computed tomography (CT). Data were used to produce
an accurate implant planning with a transfer to surgery by means of stereolithographic
drill guides. Pre-operative cone-beam CT images were subsequently matched with
post-operative ones to calculate the deviation between planned and installed implants.
Results: Placed implants (length: 10–15 mm) showed an average angular deviation of
21 (SD: 0.8, range: 0.7–4.01) as compared with the planning, while the mean linear
deviation was 1.1mm (SD: 0.7 mm, range 0.3–2.3 mm) at the hex and 2.0mm (SD:
0.7 mm, range 0.7–2.4 mm) at the tip.
Conclusions: Cone-beam images could be used for implant planning, taking into
account a maximal 41 angular and 2.4mm linear deviation at the apical tip.
Daftar copy :
No. |
Barcode |
Lokasi |
No. Rak |
Ketersediaan |
1 |
08194474 |
Ruang Referensi - Perpustakaan FKIK |
01 |
TIDAK DIPINJAMKAN |
Diproses dalam : 0.15346598625183 detik